Product Recall: Beha-Amprobe MT204-S with Test Leads (2403-0027)

Reference: 2403-0027

Product: N/A

Hazard: EN 61557-14, requires two-hand operation for test equipment when used for high voltage testing as an additional level of protection against electric shock during use. As of 1 July 2018, the shrouded test leads set initially supplied for use with the Machinery Tester, which required two-hand use, was inadvertently replaced with an alternative unshrouded test leads set that did not require two-hand operation. The inclusion of unshrouded test leads resulted in the machinery tester not being compliant with EN 61577-14 and not compliant with the essential health and safety requirements.

Corrective Action: The unshrouded product has been recalled and the machinery tester product is now shipped with shrouded test leads.

PDF: Download